Normally Ex-Stock
|
Condensate Trap WT20.5
|
WT20.5 |
|
Condensate Trap with micro filter
|
Function description:
The condensate trap is installed to the hosing directly before the gas analyser. If the gas conditioning system (cooler, peristaltic pump, dust filter etc) should fail then the semi-permeable membrane with SUN-C coating will protect the gas analyser. The membrane separates the gasses from the condensate and dust. If it becomes completely full then stream of gasses is interrupted. The condensate trap is also known as a ""police filter"". An alarm sounds via an upstream flow detector with indication ring (see example applications).
|
Technical data:
Max. water pressure: 2 Bar Max. operational gas pressure: 2 Bar Max. gas flow rate: 400 l/hr Typical flowrates 100l air/h: 10mbar Typical flowrates 400l air/h: 40mbar Diameter: approx. 7cm Length: approx. 7cm Pore-size diaphragm: <0.1 µm Maximum operative temperature: 90 C Material of construction: PP, PTFE,
|
|
|
| |
|